high contrast theme for PC-DMIS. Windows has one, but the tool bar buttons down appear so well.
Also, the EDIT WINDOW FONT colors are not very bright.

What is wrong with this 2020 version?
1. Why can't I access the most recent programs?
2. Why can't I edit nominals in the geometric tolerance tab?
I've been using this software for the past 7 years and this is one of the worst versions of the software. When you did this update, why couldn't you keep everything you had before and just add new things instead of playing with all of the old things?

RS6 scanner for all devices
Integrate RS6 scanner to all devices. to CMM for HPLs, To tracker for T-scan-Las.

FIX XACT MEASURE (DATUM SHIFT ERROR)
XACT MEASURE SHOULD UTILIZE THE PCDMIS SIZE FUNCTION WHEN EVALUATING FOR DATUM SHIFT INSTEAD OF WHAT APPEARS TO BE L^2.
WHEN MEASURING A CASTING THAT HAD AN AS CAST DATUM ID DIAMETER THAT WAS 4.4300 +/- .020, THE ACUTAL MEASURED FEATURE WAS OUT OF ROUND AND MEASURED +.0218 AND -.0235 ON SIZE, BUT WHEN USING THE XACT MEASURE IT CALCULATED AN L^2 (LEAST SQUARES) AT 4.4286 ALLOWING FOR .0186 DATUM SHIFT. WHEN THE DATUM MATING EVELOPE IS LESS THAN THE LOWEST POSSIBLE SIZE YOU DO NOT GET DATUM SHIFT. THIS ERROR NEEDS TO BE FIXED IN ORDER TO FOLLOW ASME Y14.5.

Rotatable spacer for TP20 star probes
I've never been a fan of star probes. Not to say that I haven't used them. It's more that I haven't figured out the trick to consistently and easily orient the 'star' on a TP 20 setup. Worked with a Zeiss with a Vast head for a few years. That was somewhat manageable. Seems to me that you would need some kind of hardware solution for this. Some kind of rotatable element to go in between the TP20 and your probe head. Wouldn't have to rotate 360 degrees. 90 degrees of rotation would probably be enough.

Add: Toggle to "Find" function - Constructed Features "Off"
When using the search function, a toggle box would be present to ignore searching for text in constructed features.